10 Types of Ultrasound Artifacts and How to Recognize Them

This article explaines 10 common ultrasound artifacts, their causes, and how to recognize them. Learn how probe selection and imaging features can help reduce image distortions and improve image quality.
Table of Contents

A shadow on an ultrasound image can help identify a gallstone, but it can also hide the tissue behind it. That’s the tricky part about ultrasound artifacts: they can reveal useful diagnostic information, but they can also make images harder to interpret. Knowing how to recognize them makes all the difference. In this guide, we’ll look at 10 common ultrasound artifacts, why they occur, and how to reduce their effects.

What Are Ultrasound Artifacts?

In simple terms, an ultrasound artifact is any feature on an image that does not accurately show the patient’s real anatomy. It may be a structure that appears where nothing exists, a real structure that disappears, or something that appears in the wrong position or with incorrect brightness or shape.

Different types of artifacts occur when sound waves interact with body tissues in ways that differ from the assumptions ultrasound machines use to create images, such as sound traveling in a straight line at a constant speed and echoes returning from the main beam axis. 

Although scanning techniques and machine settings can affect their appearance, most artifacts result from how sound behaves in the body rather than ultrasound machine failure. Some can even be useful, such as acoustic shadowing behind a gallstone, which is explained as follows.

What Are the Common Types of Ultrasound Artifacts?

Now that we know what an ultrasound artifact is, let’s learn about the 10 most common ones.

These common types of ultrasound artifacts can appear on different types of ultrasound machines because they come from how sound behaves in tissue, although probe design and machine features can influence their appearance.

1. Acoustic Shadowing

Let’s start with acoustic shadowing, which is a dark band or loss of echoes behind a structure in the patient’s body. This artifact happens when a structure, such as bone, a stone, or gas, strongly reflects, scatters, or attenuates the sound beam.

 

  • Gallbladder ultrasound showing a bright gallstone with a dark acoustic shadow behind itWhy it happens: When a sound wave reaches a strong reflector or highly attenuating structure in the body, much of its energy is reflected or absorbed, so very little passes through, and a shadow forms behind the structure.
  • What it looks like: A clean shadow is a black band behind a bright edge, while gas often produces a “dirty” shadow filled with scattered echoes.
  • Helpful or misleading? It can be both. Sonographers who use abdominal ultrasound machines can recognize gallstones and kidney stones based on these shadows. On the other hand, it can also hide a lesion lying behind bone, calcification, or gas.
  • How to fix it: Rescan from another angle or acoustic window so the beam passes around the obstruction. Also, use the spatial compounding feature if the ultrasound machine supports it, as it can reduce some shadowing artifacts.

2. Posterior Acoustic Enhancement

Another common ultrasound artifact is posterior acoustic enhancement. It is a brighter-than-normal area behind a fluid-filled structure, such as a cyst, the gallbladder, or the urinary bladder.

Kidney ultrasound showing a simple renal cyst with a bright band of posterior acoustic enhancement below it

  • Why it happens: Fluid attenuates less sound than the surrounding soft tissue. As a result, more acoustic energy reaches the tissue behind the fluid-filled structure, and the returning echoes appear stronger.
  • What it looks like: A bright column of tissue directly below the fluid-filled structure, in contrast with the darker tissue on either side.
  • Helpful or misleading? It can be helpful because it suggests that a lesion is filled with fluid, not solid, but it is not proof on its own. The extra brightness can also make it harder to assess real changes in the tissue behind it.
  • How to fix it: Adjust the overall gain or time gain compensation in the deeper zone when necessary to balance the image brightness, or rescan from another view. However, these adjustments change the displayed brightness rather than eliminate the physical enhancement artifact.

3. Reverberation

Reverberation is a series of evenly spaced, parallel lines or false echoes seen deeper than a strong reflector. They are created when sound bounces repeatedly between two reflective surfaces.

Ultrasound of a needle in the shoulder showing multiple equally spaced reverberation lines beneath the needle

  • Why it happens: Some echo energy is reflected again between the reflector and another surface, such as the probe face. As a result, the machine reads the delayed returns as deeper structures.
  • What it looks like: A step-ladder of lines that fade with depth. The first repeated echo can appear at about twice the depth of the real reflector in a typical reverberation pattern.
  • Helpful or misleading? It is mostly misleading. For example, in echocardiography with cardiac ultrasound probes, it can mimic a mass, thrombus, or dissection flap near a calcified wall or device lead.
  • How to fix it: Lower the gain and change the probe angle or window, since reverberation artifacts may disappear or change in appearance in other views.

4. Comet-Tail and Ring-Down

The next common ultrasound artifacts are comet-tail and ring-down. When viewing an ultrasound scan, they appear as bright streaks that extend deep to a small, highly reflective object or a collection of gas bubbles.

Two ultrasound images comparing short comet- tail artifacts from thyroid colloid with ring-down artifacts from bowel gas

  • Why it happens: Comet-tail comes from rapid reverberations between closely spaced reflectors. Ring-down is associated with the resonance of gas bubbles, which produce a continuous signal that returns to the probe.
  • What it looks like: Comet-tail appears as a narrow, tapering bright streak that fades as it goes deeper. Ring-down usually appears as a longer, continuous bright band extending behind the gas bubbles.
  • Helpful or misleading? It can be both. These artifacts can indicate gas in the bowel or gas within certain abnormal collections, but they do not confirm a gas-forming infection on their own. Comet-tail artifacts can also help identify cholesterol crystals in the gallbladder wall. On the other hand, they can obscure the tissue beneath them.
  • How to fix it: You can reposition the transducer to change how the beam meets the object, and reduce the gain if the streak obscures the structures behind it.

5. Mirror Image

A mirror image artifact is a false copy of a structure that is displayed beyond a strong, smooth reflector. This reflector can be the diaphragm, the pleura, or a large vessel wall.

Liver ultrasound showing a mirror image of the liver and a duplicated hepatic vein beyond the diaphragm
Image: Di Serafino M, et al. Common and Uncommon Errors in Emergency Ultrasound. Diagnostics 2022;12(3):631, CC BY 4.0. Cropped and annotated by Elzhen.

  • Why it happens: Part of the sound reflects off the smooth surface, travels toward the structure, and then returns to the probe along a longer path. As a result, the echo takes longer to come back, so the machine places a duplicate of the structure beyond the surface.
  • What it looks like: A fainter duplicate appears on the opposite side of the reflector, as if the reflector were a mirror. Doppler flow can also be duplicated.
  • Helpful or misleading? It is mostly misleading: it can imitate a second vessel or another structure. In veterinary ultrasound examination, a mirrored image of the liver beyond the diaphragm can mimic a structure within the thorax.
  • How to fix it: Scan from a different angle so the reflector leaves the beam path, since a mirror image may shift or disappear when the probe moves while the real structure stays in its actual position.

6. Side Lobe and Grating Lobe

Side-lobe and grating-lobe artifacts are false echoes caused by weaker ultrasound beams traveling outside the main beam. The machine may incorrectly display these echoes as if they came from the main beam.

Bladder ultrasound with false side lobe echoes in the urine, next to the same bladder appearing clear after the probe is repositioned

  • Why it happens: Besides the main beam, the probe can produce weaker beams to the sides. When these beams hit a strong reflector, the machine draws the echoes as if they came from straight ahead. Grating lobes are additional off-axis beams that can occur in array transducers because of the spacing between their elements.
  • What it looks like: It appears as a faint curved or linear band of echoes inside an otherwise echo-free area, resembling sludge, debris, or a thin membrane.
  • Helpful or misleading? It is misleading because it can imitate debris in the bladder or gallbladder. Additionally, in cardiac imaging, it can mimic a thrombus or an aortic dissection flap.
  • How to fix it: Move or tilt the probe to see whether the echoes stay, because a side-lobe or grating-lobe echo may shift or disappear while real debris stays in place. Lowering the gain or using harmonic imaging can also reduce some of these artifacts.

7. Refraction: Edge Shadowing and Double Image

Refraction artifacts occur when the beam bends at the edge of a curved structure or between tissues with different sound speeds. As a result, thin shadows can appear beside the structure, or a duplicated image may form.

Two ultrasound images showing refraction: edge shadows beside a round mass, and the aorta displayed twice on color Doppler
Left: Wu WT, et al. Diagnostics 2020;10(9):645, CC BY 4.0. Right: Nevit Dilmen, Wikimedia Commons, CC BY-SA 3.0. Annotated by Elzhen.

  • Why it happens: Sound travels at different speeds in different tissues, so a beam crossing a tilted or curved boundary at an oblique angle bends. Therefore, the machine, which assumes a straight path, draws the echoes in the wrong place.
  • What it looks like: It can appear as narrow dark lines running down from the sides of a round structure, such as a cyst (edge shadowing).
  • Helpful or misleading? It is mostly misleading, since edge shadows can hide a small lesion at the margin of a round structure, and a double image can suggest an extra structure that does not exist.
  • How to fix it: Change the probe angle or imaging window, or use spatial compounding. This way, the beam meets the boundary from a different direction.

8. Anisotropy

Anisotropy is a loss of brightness in fibrous structures such as tendons, ligaments, and nerves.

Biceps tendon ultrasound where the tendon looks falsely dark at an oblique angle and normal and bright when the beam is perpendicular
Image: Wu WT, et al. Diagnostics 2020;10(9):645, CC BY 4.0. Cropped and annotated by Elzhen.

  • Why it happens: When the beam does not meet the fibers perpendicularly, the echoes may bounce away from the probe and cause healthy tissue to look dark.
  • What it looks like: It looks like a normally bright, fibrillar tendon turns patchy and dark in the part that curves away from the probe.
  • Helpful or misleading? It is a well-known pitfall in musculoskeletal ultrasound scanning, where a healthy tendon can look like tendinosis or a partial tear.
  • How to fix it: Tilt or rock the probe until the beam is perpendicular to the fibers, and the dark area should brighten, whereas a true tear does not disappear simply by changing the probe angle.

9. Beam-Width and Slice-Thickness Artifact

Forearm ultrasound where a nerve looks blurred with the focus set too shallow and sharp when the focus is moved to nerve depth

Beam-width and slice-thickness artifacts bring echoes from tissue outside the intended scan plane into the image, so echoes seem to come from inside structures that are actually empty.

  • Why it happens: The beam is not infinitely thin and has a finite width and thickness. Therefore, the ultrasound machine combines echoes from the full width and thickness of the beam into one flat image.
  • What it looks like: It appears as fuzzy echoes inside fluid-filled structures such as the bladder or gallbladder, often forming a curved layer that resembles sludge (“pseudosludge”).
  • Helpful or misleading? It is misleading because it can imitate sludge or a mass, and with Doppler it can place flow from a nearby vessel inside a cyst or chamber.
  • How to fix it: Set the focal zone at the depth of interest and rescan in a second plane. Using a probe with a narrower beam or better elevational resolution can also help reduce this artifact.

10. Doppler Artifacts: Aliasing and Blooming

Color Doppler of a vessel showing aliasing at a low velocity scale and uniform flow color after the scale is raised
Image: Di Serafino M, et al. Diagnostics 2022;12(3):631, CC BY 4.0. Cropped and annotated by Elzhen.

Doppler artifacts are false or distorted flow signals in color or spectral Doppler.

The most common are aliasing (high velocities wrapping around the scale) and blooming (color spilling beyond the vessel).

  • Why it happens: Aliasing happens when the Doppler shift exceeds half the pulse repetition frequency (the Nyquist limit), while blooming results from color gain set too high.
  • What it looks like: Aliasing shows color wrapping around the scale or spectral peaks cut off and drawn below the baseline. Blooming makes color spill past the vessel wall.
  • Helpful or misleading? Aliasing can indicate high-velocity flow through a narrowed vessel, but it does not confirm stenosis on its own. Blooming can exaggerate vessel size or make flow appear outside the vessel. Another useful Doppler artifact is twinkling, which can help reveal small stones or calcifications.
  • How to fix it: For aliasing, raise the velocity scale (pulse repetition frequency) or shift the baseline. For blooming, lower the color gain.

Artifact or Real Finding? A Quick Checklist

Sometimes, different types of ultrasound artifacts can look very similar to a real anatomical structure or abnormality. As a sonographer, before making a medical decision based on an unusual finding, you can consider a few practical checks:

  1. Change the probe angle: Reposition or rotate the ultrasound probe to check whether the visuals change or disappear.
  2. Scan in another plane: Examine the same area from a different imaging plane to confirm that the structure is present.
  3. Compare adjacent tissues: Look at nearby structures to make sure whether the finding follows a consistent anatomical pattern.
  4. Adjust the imaging settings: Change gain, depth, focus, or frequency to check whether the appearance is related to image settings.
  5. Use another acoustic window: Rescan the same structure from a different window to see whether the finding remains.
  6. Check with Doppler when appropriate: Determine whether an apparent flow signal is real, while remembering that Doppler can also produce artifacts.
  7. Look for anatomical continuity: Follow the structure through multiple views to check whether it connects to the expected anatomy.

Paying attention to these factors can help distinguish an ultrasound artifact from a real finding. However, remember that they do not replace clinical judgment or a complete examination.

How Probe and Ultrasound Choice Affect Ultrasound Artifacts

Different types of ultrasound artifacts mainly occur because of how sound waves travel through and interact with body tissues. However, probe design and ultrasound machine features can affect how often these artifacts appear and how clearly they show up in an image.

How Probe Choice Affects Artifacts

As already explained in our article about different types of ultrasound probes, they have different frequencies, array designs, and beam characteristics, which can influence artifact appearance.

  • Frequency: Higher-frequency probes produce sharper images and better spatial resolution. However, they are not suitable for visualizing deeper structures, especially when attenuation is high.
  • Linear probes: These probes are commonly used for tendons and other superficial structures, where anisotropy is an important consideration. The probe angle must be adjusted carefully to avoid making healthy tendons look abnormal.
  • Phased array probes: These probes are commonly practical for cardiac imaging because their small footprint allows scanning between the ribs. Reverberation and side-lobe artifacts can still affect cardiac images.
  • Array design: The spacing and arrangement of transducer elements influence beam formation and can affect the production of grating lobes.

Machine Features That Help Reduce Artifacts

Modern ultrasound machines include several imaging features that can reduce certain artifacts or improve image quality.

  • Tissue harmonic imaging: This feature can reduce reverberation, side-lobe artifacts, and other forms of clutter while improving image contrast.
  • Spatial compounding: This feature combines images generated from different beam angles, which can reduce some shadowing and edge-shadowing artifacts.
  • Speckle reduction: This option reduces speckle noise and can make tissue structures easier to visualize. However, excessive processing may remove some fine details.
  • Focal zone control: Adjusting the focal zone to the depth of interest improves lateral resolution in that region, helping structures appear more clearly defined.

Choosing the right ultrasound machine goes beyond comparing image quality or technical configurations. The appropriate system depends on the examination type, clinical applications, transducer requirements, and imaging features needed for daily work.

At Elzhen, we help hospitals, clinics, and medical professionals choose the right ultrasound machines and transducers based on their clinical needs and budget. Whether you are looking for a refurbished ultrasound machine for general imaging, cardiac examinations, or specialized applications, our team can help you find a suitable solution.

Medical References

FAQ
Ultrasound artifacts are image distortions that can make anatomical structures look different from their actual appearance.
Ultrasound artifacts are created from sound waves reflecting, bending, or losing energy as they travel through different tissues. Transducer characteristics, scanning angles, and imaging settings can also influence their appearance.
Sonographers can reduce certain artifacts by adjusting the transducer angle, selecting an appropriate frequency, changing the imaging window, and using machines that feature harmonic imaging or spatial compounding.
Artifacts that hide important anatomy or imitate abnormal findings can be particularly problematic. Acoustic shadowing, reverberation, and side-lobe artifacts may interfere with image interpretation if they are not recognized.
Not necessarily. Most ultrasound artifacts result from how sound waves interact with body tissues, although transducer design, equipment condition, and image-processing features can influence their appearance.